Stimulus state, evolution, fluctuation, and static modifiers

Stimulus modifiers describe reproducible response to alerting stimulation; temporal modifiers describe frequency, morphology, or location change.

Criteria & assessment

stimulus state
SI
Reproducibly brought about or exacerbated by alerting stimulus in a less-stimulated state, with or without clinical alerting; may also occur spontaneously.
ST
Reproducibly terminated or attenuated by alerting stimulus in a less-stimulated state, with or without clinical alerting; may also self-terminate.
spontaneous
Never clearly induced, exacerbated, improved, or terminated by stimulation.
unknown
Unclear or untested.
stimulus type required
  • auditory
  • light tactile
  • patient care/other nonnoxious
  • noxious
evolving
minimum
At least 2 unequivocal sequential changes in the same category.
frequency
Two consecutive changes in one direction, each at least 0.5 Hz; each frequency persists at least 3 cycles.
morphology
Two consecutive novel morphologies; each morphology, including transition forms, persists at least 3 cycles.
location
Sequential spread into or out of at least 2 standard 10-20 electrode locations; involvement of each additional electrode persists at least 3 cycles.
interstep gap
Reach criteria without the changing feature remaining unchanged for 5 or more continuous minutes.
amplitude
Voltage/amplitude change alone is neither evolution nor a new morphology.
fluctuating
minimum
At least 3 changes, no more than 1 minute apart, not qualifying as evolution.
frequency
Each change at least 0.5 Hz; each frequency persists at least 3 cycles.
morphology
Each morphology persists at least 3 cycles.
location
Each change at least 1 standard interelectrode distance; each location persists at least 3 cycles.
amplitude
Voltage/amplitude change alone is not fluctuation.
static
Neither evolving nor fluctuating.
additional reporting
If evolving or fluctuating in frequency, specify minimum and maximum frequency. For nongeneralized patterns specify spread as none, unilateral, or bilateral.
Generation & measurement notes
  • Use an event sequence rather than a single classifier score so cycle persistence, direction, and interstep gap can be audited.

Editorial draft · clinical review pending